The THS-100 is a fully automated headspace sampler featuring a 100-position sampling system. It is widely used in fields such as healthcare, environmental protection, petrochemicals, the food industry, and pharmaceuticals.

Features:
Standard 100-position sample tray
Equipped with a 7-inch large touch screen
Utilizes an XYZ three-axis motion platform
Automatic leak detection for sample vials and tubing

Introduction

Parameters

Product Introduction

The THS-100 is a high-throughput, automated headspace sampler equipped with a 100-position sample tray, designed specifically for use with Gas Chromatography (GC) and Gas Chromatography-Mass Spectrometry (GC-MS) systems. Featuring an XYZ-axis motion platform and Electronic Pressure Control (EPC) technology, the instrument supports simultaneous multi-sample heating and agitation, significantly enhancing analytical efficiency. The system complies with multiple national standards for environmental monitoring and health inspection—including HJ 736, HJ 741, and GA/T 842—and is widely used in fields such as environmental soil and water analysis, food safety testing, blood alcohol determination, and pharmaceutical R&D.

Key Technical Features


High-throughput automatic sampling: Standard 100-position sample tray supporting 20 mL and 10 mL headspace vials, meeting the needs for continuous analysis of large sample batches.Intuitive user interface: Equipped with a 7-inch touchscreen displaying real-time instrument operation graphics, module parameters, and individual sample status for easy operation.

Efficient heating and transfer system: Utilizes a multi-position heating tray for overlapping sample heating, combined with agitation at heating positions to shorten equilibration time; the sampling needle and transfer lines feature an inert, “cold-spot-free” design to prevent sample loss due to condensation.

Broad compatibility: Compatible with various mainstream brands of Gas Chromatographs (GC) and Gas Chromatography-Mass Spectrometers (GC-MS).

Intelligent control and safety monitoring: Built-in Electronic Pressure Controller (EPC) for precise carrier gas pressure regulation; includes automatic leak detection for vials and transfer lines to ensure system integrity.

Data traceability and management: Logs complete analysis methods and conditions, facilitating experimental data tracking and compliance management.

THS-100 Headspace Sampler

Sample No. Data File Retention Time Area Height Concentration (ppm)
1 12.gcd 3.574 54347 24106 50
2 13.gcd 3.573 54738 24354 50
3 14.gcd 3.573 54863 24529 50
4 15.gcd 3.573 54469 24295 50
5 16.gcd 3.573 53960 24039 50
6 17.gcd 3.573 54836 24468 50
7 18.gcd 3.573 53874 23998 50
8 19.gcd 3.573 53785 23943 50
9 20.gcd 3.573 53475 23875 50
  Average 3.573 54261 24178 50
  %RSD 0.01% 0.93% 0.99%  

 

Instrument Specifications
Sample positions 100 positions, 20 mL headspace vials (10 mL optional)
Gas path pressure rating >0.4 MPa
Synchronization signal output Two channels of 1–2 second switch signals
Instrument dimensions Height × Width × Length: 460 mm × 605 mm × 860 mm